Steam autoclave troubleshooting: wet packs, failed indicators and long cycles

This guide covers common symptoms and the questions to ask before calling for service. Always follow the manufacturer's manual, and stop using a sterilizer when a safety or sterility problem is suspected.

SymptomCommon causes to check
Wet packs after the cycleHeavy metal mass, dense wraps, crowded loads, skipped or short drying, opening the door too soon
Chemical indicator did not changeSteam did not reach the item: dense packaging, trapped air, wrong cycle, overloaded chamber
Failed Bowie-Dick testAir leaks, door gasket problems, vacuum system faults or poor steam supply. Stop using the unit and call for service
Positive biological indicatorProcess failure or a testing error. Quarantine, retest and follow your recall procedure
Cycles taking longer than usualLarge or dense loads, heating problems, water supply issues
Stains or spots on instrumentsWater quality, residues from cleaning or detergents

Daily care that prevents problems

When to call for service

Call when a Bowie-Dick or biological test fails without an obvious loading cause, when error codes repeat, or when the unit leaks steam or water. After a repair, test before returning the sterilizer to routine use, as the monitoring guide describes.